ENDOTAK RELIANCE G
Report
- Report Number
- 2124215-2025-08269
- Event Type
- Malfunction
- Date Received
- February 13, 2025
- Date of Event
- July 28, 2023
- Report Date
- March 24, 2026
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NVY
- UDI-DI
- 00802526531262
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IF PERTINENT INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IF PERTINENT INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD EXHIBITED HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS, WITH RECENT MEASUREMENTS IN THE 132-133 OHMS RANGE. IT WAS NOTED THAT THIS LEAD HAD A DELIVERED SHOCK IMPEDANCE OF 88 OHMS FROM 2021. TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS AND PROGRAMMING CONSIDERATIONS. THIS RV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS OR INTERVENTIONS WERE REPORTED AT THIS TIME.
IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD EXHIBITED HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS, WITH RECENT MEASUREMENTS IN THE 132-133 OHMS RANGE. IT WAS NOTED THAT THIS LEAD HAD A DELIVERED SHOCK IMPEDANCE OF 88 OHMS FROM 2021. TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS AND PROGRAMMING CONSIDERATIONS. THIS RV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS OR INTERVENTIONS WERE REPORTED AT THIS TIME. ADDITIONAL INFORMATION RECEIVED REPORTED THAT RV LEAD CONTINUED TO EXHIBIT HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS, WITH A MEASUREMENT AS HIGH AS 148 OHMS IN NOVEMBER 2025. TECHNICAL SERVICES (TS) DISCUSSED PROGRAMMING CONSIDERATIONS AND NEXT STEPS, INCLUDING POSSIBLE LEAD REVISION. THIS RV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS OR INTERVENTIONS WERE REPORTED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1490671 | ENDOTAK RELIANCE G | PERMANENT DEFIBRILLATOR ELECTRODES | NVY | BOSTON SCIENTIFIC CORPORATION | 0295 | 419703 | 00802526531262 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Female |