FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 21379521 · Received February 13, 2025

Report

Report Number
1213809-2025-00091
Event Type
Malfunction
Date Received
February 13, 2025
Date of Event
January 21, 2025
Report Date
April 13, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4)- SUPPLEMENTAL MDR - OTHER TWO HUNDRED SAMPLES FROM BATCH 4263689 AND TWO HUNDRED SAMPLES FROM BATCH 4263696 WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. ACCORDING TO VERBATIM, THE COMPLAINT APPEARS TO BE FOR THE ABSENCE OF STOP RING ON THE PLUNGER ROD. THIS PRODUCT DOES NOT HAVE A STOP RING BY DESIGN ACCORDING TO ITS PRODUCT SPECIFICATION. BATCHES 3285600, 4033569, 4158490, 4263689 AND 4263689 ARE CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

D.4. OTHER LOT NUMBERS INCLUDE 4033569, 4158490, 4263689 AND 4263696. OTHER EXPIRATION DATES INCLUDE 2029-01-31, 2029-05-31 AND 2029-08-31. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. H.4. OTHER LOT NUMBER DEVICE MANUFACTURE DATES ARE 2024-02-14, 2024-07-15 AND 2024-09-19.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK PLUNGER "DOES COME OUT QUITE EASILY IF THE PLUNGER IS PULLED". THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NO RESISTANCE AT THE END OF TRAVEL. IT DOES COME OUT QUITE EASILY IF THE PLUNGER IS PULLED WITHOUT PAYING ATTENTION WHEN USING THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1125840 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 3285600 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown