FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 21377346 · Received February 13, 2025

Report

Report Number
1722139-2025-00069
Event Type
Malfunction
Date Received
February 13, 2025
Date of Event
January 15, 2025
Report Date
January 15, 2025
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
UDI-DI
10814844000075
PMA / PMN Number
K031199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED TO MMD FOR EVALUATION. A DHR REVIEW WAS COMPLETED AND NO NON-CONFORMANCES WERE FOUND. BECAUSE THE DEVICE WAS NOT RETURNED TO MMD FOR EVALUATION, AN INVESTIGATION COULD NOT BE COMPLETED. THIS REPORT WILL BE UPDATED IF THE DEVICE IS RETURNED TO MMDG.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO MMD FOR EVALUATION. A DHR REVIEW WAS COMPLETED AND NO NON-CONFORMANCES WERE FOUND. WHEN THE DEVICE WAS RETURNED TO MMD FOR INVESTIGATION, IT DID OVER INFUSE, HOWEVER, IT WAS STILL WITHIN THE VOLUMETRIC RANGE THAT MMD WOULD CONSIDER NOT REPORTABLE. BASED ON THIS INFORMATION, NO MDR WOULD HAVE BEEN REQUIRED.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THAT THE PUMP UNDER INFUSED AT A RATE THE MMD WOULD CONSIDER REPORTABLE. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE PATIENT HAD NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THE COMPLAINT. NO ADDITIONAL INFORMATION WAS PROVIDED. (B)(4)

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THAT THE PUMP UNDER INFUSED AT A RATE THE MMD WOULD CONSIDER REPORTABLE. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE PATIENT HAD NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THE COMPLAINT. NO ADDITIONAL INFORMATION WAS PROVIDED. [COMPLAINT- (B)(4)].

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100739 ENTERALITE INFINITY ENTERAL FEEDING PUMP ENTERAL INFUSION PUMP LZH MOOG MEDICAL DEVICES GROUP INFKIT2 N/A 10814844000075

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose