ENTERALITE INFINITY ENTERAL FEEDING PUMP
Report
- Report Number
- 1722139-2025-00069
- Event Type
- Malfunction
- Date Received
- February 13, 2025
- Date of Event
- January 15, 2025
- Report Date
- January 15, 2025
- Manufacturer
- MOOG MEDICAL DEVICES GROUP
- Product Code
- LZH
- UDI-DI
- 10814844000075
- PMA / PMN Number
- K031199
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
THE DEVICE WAS NOT RETURNED TO MMD FOR EVALUATION. A DHR REVIEW WAS COMPLETED AND NO NON-CONFORMANCES WERE FOUND. BECAUSE THE DEVICE WAS NOT RETURNED TO MMD FOR EVALUATION, AN INVESTIGATION COULD NOT BE COMPLETED. THIS REPORT WILL BE UPDATED IF THE DEVICE IS RETURNED TO MMDG.
THE DEVICE WAS RETURNED TO MMD FOR EVALUATION. A DHR REVIEW WAS COMPLETED AND NO NON-CONFORMANCES WERE FOUND. WHEN THE DEVICE WAS RETURNED TO MMD FOR INVESTIGATION, IT DID OVER INFUSE, HOWEVER, IT WAS STILL WITHIN THE VOLUMETRIC RANGE THAT MMD WOULD CONSIDER NOT REPORTABLE. BASED ON THIS INFORMATION, NO MDR WOULD HAVE BEEN REQUIRED.
THE INITIAL REPORTER STATED THAT THE PUMP UNDER INFUSED AT A RATE THE MMD WOULD CONSIDER REPORTABLE. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE PATIENT HAD NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THE COMPLAINT. NO ADDITIONAL INFORMATION WAS PROVIDED. (B)(4)
THE INITIAL REPORTER STATED THAT THE PUMP UNDER INFUSED AT A RATE THE MMD WOULD CONSIDER REPORTABLE. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE PATIENT HAD NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THE COMPLAINT. NO ADDITIONAL INFORMATION WAS PROVIDED. [COMPLAINT- (B)(4)].
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100739 | ENTERALITE INFINITY ENTERAL FEEDING PUMP | ENTERAL INFUSION PUMP | LZH | MOOG MEDICAL DEVICES GROUP | INFKIT2 | N/A | 10814844000075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |