FDA Adverse Event Malfunction Summary report: N

I-STAT ACTK (KAOLIN) CARTRIDGE

MDR report key: 2137375 · Received May 31, 2011

Report

Report Number
2245578-2011-00079
Event Type
Malfunction
Date Received
May 31, 2011
Date of Event
May 5, 2011
Report Date
May 30, 2011
Manufacturer
ABBOTT POINT OF CARE CANADA LTD.
Product Code
JBP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED A DELAY IN OBTAINING RESULTS WITH AN I-STAT ACT KAOLIN CARTRIDGE. THE CARTRIDGE WAS RUNNING AND THE ANALYZER TIMED OUT AFTER 15 MINUTES ON THREE OCCASIONS. ON (B)(6) 2011, A PATIENT WAS UNDERGOING AN ELECTROPHYSIOLOGY PROCEDURE AND ACT WAS BEING PERFORMED EVERY 15 MINUTES. THE FOLLOWING RESULTS (IN SECONDS) WERE OBTAINED: 227, 277, >1000, 340, 315, 340, >1000, 350, 365, 365, 340, >1000.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT ACTK (KAOLIN) CARTRIDGE ACTK CARTRIDGE JBP ABBOTT POINT OF CARE CANADA LTD. NA S11082B

Patients

Seq Age Sex Outcome Treatment
1 63 YR