AZURE XT DR MRI SURESCAN
Report
- Report Number
- 9614453-2025-00626
- Event Type
- Malfunction
- Date Received
- February 13, 2025
- Date of Event
- December 12, 2024
- Report Date
- February 13, 2025
- Manufacturer
- MEDTRONIC EUROPE SARL
- Product Code
- NVZ
- UDI-DI
- 00643169871939
- PMA / PMN Number
- P980035
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED THAT THERE WAS ELECTROMAGNETIC INTERFERENCE (EMI)/OVERSENSING NOTED ON STORED ELECTROGRAMS (EGM). THE IMPLANTABLE PU LSE GENERATOR (IPG) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126618 | AZURE XT DR MRI SURESCAN | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC EUROPE SARL | W1DR01 | 00643169871939 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Male | 383069 LEAD, 407652 LEAD |