FDA Adverse Event Injury Summary report: N

OMNIPOD 5 POD

MDR report key: 21366341 · Received February 12, 2025

Report

Report Number
3004464228-2025-06358
Event Type
Injury
Date Received
February 12, 2025
Date of Event
December 26, 2024
Report Date
February 12, 2025
Manufacturer
INSULET CORPORATION
Product Code
QFG
UDI-DI
10385083000213
PMA / PMN Number
K203768
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO DETERMINE IF ANY PRODUCT CONDITION COULD HAVE CONTRIBUTED TO THE REPORTED HOSPITALIZATION AND DIABETIC KETOACIDOSIS. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS. *PLEASE NOTE, THAT SECTION D IS CAPTURING THE DEVICE IDENTIFIERS AS REPORTED BY THE COMPLAINANT.   THIS MAY NOT ALIGN TO THE DEVICE CONFIGURATION REPORTED IN H11, AS THIS DATA IS PULLED FROM OUR CLOUD BASED ON THE REPORTED DATE OF EVENT. CLOUD - LOCKED DOWN/SMARTPHONE: LOCKDOWN CLOUD - OMNIPOD 5 SOFTWARE APP VERSION: 3.1.1. CLOUD - SMARTPHONE OPERATING SYSTEM: N5004L-AM-Q-MV01602-06-01.06. CLOUD - SMARTPHONE HARDWARE: N5004L. CLOUD - CGM SENSOR TYPE: G6.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT HAD BEEN HOSPITALIZED WITH DIABETIC KETOACIDOSIS (DKA). THE PATIENT REPORTS HAVING A LOSS OF CONSCIOUSNESS WHILE WEARING A POD. ONCE AT THE HOSPITAL THE PATIENT WAS PLACED IN THE INTENSIVE CARE UNIT. TREATMENT INFORMATION IS UNAVAILABLE. THE PATIENT WAS IN THE HOSPITAL INTENSIVE CARE UNIT FOR 26 DAYS. THE PATIENTS POD WILL NOT BE RETURNED AS IT HAS BEEN DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1042590 OMNIPOD 5 POD ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP QFG INSULET CORPORATION PT-000536 PH1K11162351 10385083000213

Patients

Seq Age Sex Outcome Treatment
1 69 YR Female Hospitalization