FDA Adverse Event Malfunction Summary report: N

2.8MM QUICK RELEASE DRILL

MDR report key: 21357485 · Received February 11, 2025

Report

Report Number
3025141-2025-00060
Event Type
Malfunction
Date Received
February 11, 2025
Date of Event
January 28, 2025
Report Date
February 11, 2025
Manufacturer
ACUMED, LLC
Product Code
HTW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURING AND INSPECTION RECORDS WERE REVIEWED, AND NO ANOMALIES WERE FOUND. HOWEVER, THE REPORTED 1.5MM HEX DRIVER TIP, LOCKING GROOVE (PART NUMBER 80-0728, BATCH NUMBER 599700), 2.8MM X 5" QUICK RELEASE DRILL (PART NUMBER MS-DC28, BATCH NUMBER 447604), AND 2.8MM QUICK RELEASE DRILL (PART NUMBER 80-0387, BATCH NUMBER 594442) WERE NOT RETURNED FOR EVALUATION. ADDITIONALLY, THE REPORTED SCREW WAS NOT RETURNED FOR EVALUATION. MANUFACTURING AND INSPECTION RECORDS COULD NOT BE REVIEWED AS THE MODEL NUMBER AND BATCH/LOT NUMBER OF THE SCREW IS UNKNOWN.

Description of Event or Problem · 0

(REPORT 4 OF 4) IT WAS REPORTED DURING SURGERY WHILE THE SURGEON WAS INSERTING THE LOCKING SCREW THE DRIVER TIP BROKE. THE SURGEON WAS ABLE TO REMOVE THE BROKEN PIECE AND THEREFORE IT WAS NOT LEFT IN THE PATIENT'S BODY. IT WAS ALSO REPORTED THE DRILL PENETRATED BONE AND SOFT TISSUE WHILE DRILLING WAS PERFORMED. THE DRILLS CONTACTED THE OPERATING TABLE RESULTING IN ONE DRILL BENDING AND ANOTHER BREAKING. THE SURGERY WAS COMPLETED WITH REPLACEMENT DEVICES AFTER A 10-MINUTE DELAY. NO OTHER PATIENT CONSEQUENCES WERE REPORTED. THERE ARE 4 RELATED REPORT NUMBERS FOR THIS EVENT 3025141-2025-00057.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
966269 2.8MM QUICK RELEASE DRILL BIT, DRILL HTW ACUMED, LLC 80-0387 594442

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female