SWAN GANZ TRUE SIZE
Report
- Report Number
- 2015691-2025-00927
- Event Type
- Malfunction
- Date Received
- February 11, 2025
- Report Date
- March 21, 2025
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DQO
- UDI-DI
- 00690103000245
- PMA / PMN Number
- K160084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THE DEVICE WAS REQUESTED TO BE RETURNED, BUT DID NOT ARRIVE FOR PRODUCT EVALUATION. THEREFORE, A ROOT CAUSE COULD NOT BE DETERMINED. THE DEVICE HISTORY RECORD REVIEW WAS COMPLETED AND ALL MANUFACTURING INSPECTIONS PASSED WITH NO NON-CONFORMANCES. AN ENGINEERING EVALUATION WAS INITIATED TO ASSESS FOR ANY MANUFACTURING RELATED PROCESSES WHICH COULD BE CORRELATED TO THE ISSUE. AS PART OF THE MANUFACTURING PROCESS 100 PERCENT OF THE BALLOONS GO THROUGH A BALLOON INSPECTION. BASED ON THE AVAILABLE INFORMATION, THERE IS NO EVIDENCE THAT SUPPORTS OR CONFIRMS THE FAILURE MODE IS ASSOCIATED WITH A MANUFACTURING OR DESIGN DEFECT. IT IS NOT KNOWN IF PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THE MONTHLY REVIEW.
THE DEVICE HAS BEEN REQUESTED TO BE RETURNED BUT HAS NOT YET ARRIVED. ONCE IT HAS ARRIVED AND THE PRODUCT EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL SUBMISSION WILL BE SENT. THE DEVICE HISTORY RECORD REVIEW IS PENDING. ONCE IT HAS BEEN COMPLETED THE RESULTS WILL BE SENT IN A SUPPLEMENTAL SUBMISSION. ADDITIONAL PRODUCT CODES, DYG, DQE, KRA.
AS REPORTED, DURING SET UP, THERE WAS AN ISSUE WITH A SWAN GANZ CATHETER BALLOON. DURING TESTING OF THE BALLOON, IT WOULD NOT PROPERLY INFLATE. IT WAS UNDERINFLATED AFTER TESTING AND THEN IT WOULD NOT DEFLATE. THERE WAS NO PATIENT HARM OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1753455 | SWAN GANZ TRUE SIZE | SWAN GANZ TRUE THERMODILUTION CATHETER | DQO | EDWARDS LIFESCIENCES | 096F6P | 65906463 | 00690103000245 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |