FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 2135360 · Received May 17, 2011

Report

Report Number
1824206-2011-02728
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 18, 2011
Report Date
April 18, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD TECHNICIAN FOUND THAT THE BLOWER BOX HAD HUNG UP ON J CHANNEL WHEN BED WAS LOWERING, CONTACTING THE J CHANNEL WHEN THE BED WAS ALMOST ALL THE WAY DOWN. HE REPOSITIONED THE BLOWER BRACKET WHERE IT WOULDN'T CONTACT J FRAME TO REPAIR THE BED.

Description of Event or Problem · 1

THE HOSPITAL STAFF STATED BED HI/LOW FELL TWO INCHES FOR NO REASON. THE BED WAS ALMOST ALL THE WAY DOWN AND THE NURSE SAID IT JUST FELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1 UNK