FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 2135360
·
Received May 17, 2011
Report
- Report Number
- 1824206-2011-02728
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 18, 2011
- Report Date
- April 18, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FIELD TECHNICIAN FOUND THAT THE BLOWER BOX HAD HUNG UP ON J CHANNEL WHEN BED WAS LOWERING, CONTACTING THE J CHANNEL WHEN THE BED WAS ALMOST ALL THE WAY DOWN. HE REPOSITIONED THE BLOWER BRACKET WHERE IT WOULDN'T CONTACT J FRAME TO REPAIR THE BED.
Description of Event or Problem · 1
THE HOSPITAL STAFF STATED BED HI/LOW FELL TWO INCHES FOR NO REASON. THE BED WAS ALMOST ALL THE WAY DOWN AND THE NURSE SAID IT JUST FELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |