FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2135347
·
Received June 21, 2011
Report
- Report Number
- 2531779-2011-04308
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Report Date
- May 20, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON 06/06/2011. (B)(6). RECALL 2531779-03/24/2010-003-R. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/06/2011 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS NOT DUPLICATED DURING TESTING.
Description of Event or Problem · 1
PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED PARTIALLY DISLODGED FORCE SENSOR PINS. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.