BLANKETROL III
Report
- Report Number
- 1516825-2011-00007
- Event Type
- Other
- Date Received
- June 11, 2011
- Date of Event
- December 13, 2010
- Report Date
- May 13, 2011
- Manufacturer
- CINCINNATI SUB-ZERO PRODUCTS, INC.
- Product Code
- DWJ
- PMA / PMN Number
- K101589
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
THIS COMPLAINT WAS ORIGINALLY RECEIVED ON (B)(4) 2011 ((B)(4)). AT THE TIME, CSZ DECIDED THAT THIS COMPLAINT DOES NOT QUALIFY TO BE REPORTABLE UNDER 803, AND HENCE NO MDR WAS SUBMITTED. ON (B)(6) 2011, CSZ RECEIVED A USER FACILITY REPORT (UFR) FROM FDA AND DECIDED TO SUBMIT A RESPONSE VIA MDR FORM 3500A. AT THE TIME OF ORIGINAL COMPLAINT REPORTING IN (B)(6) 2011, (B)(6), MET WITH (B)(6) ON (B)(6) 2011, AT THE FACILITY. (B)(6) REPORTED THAT THE ALLEGED "COLD BURN" WAS ACTUALLY JUST MARKING ON THE SKIN. THE PT APPARENTLY HAD MARKINGS ACROSS THE FRONT AND BACKSIDE OF THE HEAD AND TWO IMPRESSIONS ON THE BACK. THE HEAD MARKINGS WERE A RESULT OF THE HEAD WRAP BEING APPLIED TOO TIGHT AND THE MARKINGS ON THE BACK WERE A RESULT OF BEING POSITIONED DIRECTLY ON THE TWO HOSE CONNECTIONS ON THE PT VEST. THIS ISSUE IS AN APPARENT USER ERROR. (B)(6) HAS PROVIDED IN-SERVICE TO THE STAFF 3 TIMES PREVIOUSLY AND HAS OFFERED TO CONDUCT ADDITIONAL IN-SERVICING. ADDITIONALLY, CSZ CLINICAL CONSULTANT (B)(6), HAS ALSO IN-SERVICED THE STAFF AT THE FACILITY. AS A RESULT OF RECEIVING THIS UFR, CSZ, ON (B)(6) 2011, REQUESTED ADDITIONAL INFO FROM THE FACILITY DIRECTOR OF RISK MANAGEMENT BUT HAS NOT RECEIVED ANY INFO TO DATE. A FOLLOW UP REMINDER HAS BEEN SENT TO THE DIRECTOR OF RISK MANAGEMENT, AS SOON AS THE INFO IS RECEIVED AND FOUND TO BE DIFFERENT THAN WE HAD EARLIER, WE WILL SUBMIT SUPPLEMENTAL MDR ACCORDINGLY.
CARDIAC ARREST PT WHO MET THE CRITERIA TO BE A CODE ICE SUSTAINED SKIN BREAKDOWN (FROST BITE TYPE BURN) FROM THE TUBES OF THE CINCINNATI MODEL BLANKETROL III UNIT TO THE BACK OF HIS HEAD, RIGHT THIGH, AND RIGHT MID BACK. THE COOLING WATER WAS NOT FLOWING THROUGH THE THERMASUIT UNTIL THE PT ARRIVED IN THE CRITICAL CARE UNIT (CVICU) FROM THE ED SO THE STAFF COULD NOT GET A TEMPERATURE UNTIL THE PT ARRIVED IN THE CVICU. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BLANKETROL III | THERMAL REGULATING SYSTEM | DWJ | CINCINNATI SUB-ZERO PRODUCTS, INC. | 233 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Other |