SERVO-S BASE UNIT
Report
- Report Number
- 8010042-2025-0000223
- Event Type
- Malfunction
- Date Received
- February 10, 2025
- Date of Event
- February 3, 2025
- Report Date
- February 10, 2025
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K123149
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON-SITE INVESTIGATION WAS PERFORMED BY FIELD SERVICE ENGINEER. THE CLAIMED ISSUE WAS REPRODUCED DURING TROUBLESHOOTING. ACCORDING TO INFORMATION RECEIVED IN THE SERVICE REPORT, THE DEVICE FAILED INTERNAL LEAKAGE TEST, PRESSURE TRANSDUCER TEST AND PATIENT CIRCUIT TEST DURING PRE-USE CHECK. REPLACEMENT OF THE PRESSURE TRANSDUCER PRINTED CIRCUIT BOARD (PCB) SOLVED THE ISSUE. THE VENTILATOR PASSED ALL FUNCTIONAL AND SAFETY TESTS AND WAS CLEARED FOR CLINICAL USE. THE DEFECTIVE PART WAS NOT RETURNED FOR INVESTIGATION, DESPITE REQUEST. PRESSURE TRANSDUCER PCB MEASURES THE PRESSURE, CONVEYED VIA THE PRESSURE TUBE CONNECTED TO THIS BLOCK BY ITS DIFFERENTIAL PRESSURE TRANSDUCER. WITH DIFFERENTIAL REFERENCE TO THE AMBIENT PRESSURE, THE OUTPUT SIGNAL IS PROPORTIONAL TO THE MEASURED PRESSURE THUS GIVING A LINEAR MEASUREMENT IN THE RANGE -40 CMH2O TO +160 CMH2O. THE CAUSE OF THE CLAIMED ISSUE IN THIS CUSTOMER PRODUCT COMPLAINT HAS BEEN DETERMINED. THE ISSUE WAS RELATED TO PRESSURE TRANSDUCER PCB.
THE UDI INFORMATION IS NOT AVAILABLE SINCE THE DEVICE WAS MANUFACTURED BEFORE 2015.
MANUFACTURER'S REF. #: (B)(4).
IT WAS REPORTED THAT THE VENTILATOR FAILED MULTIPLE TESTS DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: 1220589.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 835755 | SERVO-S BASE UNIT | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB | 6640440 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |