FDA Adverse Event Injury Summary report: N

INSYNC III MARQUIS

MDR report key: 2134026 · Received June 20, 2011

Report

Report Number
6000094-2011-01052
Event Type
Injury
Date Received
June 20, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
PMA / PMN Number
P010031/S018
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED IN-PART ON DEVICE ANALYSIS. THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED IN-PART ON DEVICE ANALYSIS. THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYSIS FOUND NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED DURING THE DEVICE CHANGE OUT THE VENTRICULAR LEAD THRESHOLD WAS POOR. THE LEAD WAS REPLACED. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. UPON FURTHER ANALYSIS, THE DEVICE DID NOT TEST OUT OF SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED DURING THE DEVICE CHANGE OUT THE VENTRICULAR LEAD THRESHOLD WAS POOR. THE LEAD WAS REPLACED. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSYNC III MARQUIS IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC S.A. 7279 ASKU

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD