FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2134016 · Received June 20, 2011

Report

Report Number
2649622-2011-09994
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
April 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE PHYSICIAN ATTEMPTED TO EXTEND THE HELIX BEFORE IMPLANTATION AND QUESTIONED THE HELIX MECHANISM OF THE LEAD. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Other