FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2133987 · Received June 20, 2011

Report

Report Number
2649622-2011-09981
Event Type
Injury
Date Received
June 20, 2011
Date of Event
April 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM, THERE WAS APPARENT EXPLANT DAMAGE AND THERE WAS A NON-ELECTRICAL MISCELLANEOUS FINDING ON THE SLEEVE HEAD (GUIDE TOOTH DAMAGED).

Description of Event or Problem · 1

IT WAS REPORTED THAT BOTH THE ATRIAL AND RIGHT VENTRICULAR LEADS HAD FAILED AND WERE DISLODGED. THE ATRIAL LEAD WAS PACING THE VENTRICLE. DURING THE IMPLANT PROCEDURE THE PREVIOUS DAY, THE LEADS HAD SENSING DIFFICULTIES AND IT TOOK TOO MANY TURNS TO EXTEND THE HELIXES ON THE LEADS. THE PHYSICIAN EXPRESSED DISSATISFACTION WITH THE LEADS. BOTH LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 89 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR