FDA Adverse Event Malfunction Summary report: N

MAVERICK2 BALLOON CATHETER

MDR report key: 2133967 · Received June 20, 2011

Report

Report Number
2134265-2011-02314
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
May 25, 2011
Report Date
May 26, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 3.75MM X 15MM MAVERICK 2 MONORAIL BALLOON CATHETER WAS ADVANCED TO THE UNSPECIFIED TARGET LESION. ON THE FIRST INFLATION THE BALLOON REACHED 10 ATMS AND RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AND THE PATIENT'S CURRENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAVERICK2 BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493892815370 13148637

Patients

Seq Age Sex Outcome Treatment
1