FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2133781 · Received May 12, 2011

Report

Report Number
1824206-2011-02683
Event Type
Malfunction
Date Received
May 12, 2011
Date of Event
April 20, 2011
Report Date
April 20, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BED WOULD GO IN FORWARD DRIVE BUT NOT REVERSE DRIVE. THE TECHNICIAN ISOLATED THE ISSUE TO THE LEFT TRANSPORT HANDLE. HE REPLACED THE TRANSPORT HANDLE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO REC'D INDICATES WHEN THE INTELIDRIVE HANDLES ARE PULLED BACK, THE BED DRIVES FORWARD. NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1