WANDA PTA BALLOON DILATATION CATHETER
Report
- Report Number
- 2134265-2011-02264
- Event Type
- Malfunction
- Date Received
- June 20, 2011
- Date of Event
- May 23, 2011
- Report Date
- May 23, 2011
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- LIT
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4).
DEVICE EVALUATED BY MANUFACTURER: VISUAL AND TACTILE EXAMINATION OF THE RETURNED DEVICE REVEALED NO DAMAGE WAS NOTED TO THE SHAFT OF THE DEVICE. THE BALLOON WAS FOLDED AND WRAPPED FULLY AROUND THE SHAFT OF THE DEVICE. THE DEVICE WAS ATTACHED TO AN ENCORE INFLATION UNIT AND AN ATTEMPT WAS MADE TO INFLATE THE BALLOON WHEN A LEAK WAS NOTED. VISUAL AND MICROSCOPIC EXAMINATION OF THE BALLOON OBSERVED A LESS THAN 1MM LONGITUDINAL TEAR IN THE BALLOON MATERIAL APPROXIMATELY 22MM PROXIMAL TO THE PROXIMAL EDGE OF THE DISTAL MARKERBAND. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE NON-CALCIFIED CEPHALIC VEIN. ON THE FIRST INFLATION OF THE 6.0X40MM WANDA PTA BALLOON IT WAS INFLATED TO 12 ATM WITH AN ENCORE 26 INFLATION DEVICE AND THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE NON-CALCIFIED CEPHALIC VEIN. ON THE FIRST INFLATION OF THE 6.0X40MM WANDA PTA BALLOON IT WAS INFLATED TO 12 ATM WITH AN ENCORE 26 INFLATION DEVICE AND THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WANDA PTA BALLOON DILATATION CATHETER | CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL | LIT | BOSTON SCIENTIFIC - GALWAY | H965SCH505100 | 14066228 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |