FDA Adverse Event Malfunction Summary report: N

WANDA PTA BALLOON DILATATION CATHETER

MDR report key: 2133770 · Received June 20, 2011

Report

Report Number
2134265-2011-02264
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
May 23, 2011
Report Date
May 23, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LIT
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: VISUAL AND TACTILE EXAMINATION OF THE RETURNED DEVICE REVEALED NO DAMAGE WAS NOTED TO THE SHAFT OF THE DEVICE. THE BALLOON WAS FOLDED AND WRAPPED FULLY AROUND THE SHAFT OF THE DEVICE. THE DEVICE WAS ATTACHED TO AN ENCORE INFLATION UNIT AND AN ATTEMPT WAS MADE TO INFLATE THE BALLOON WHEN A LEAK WAS NOTED. VISUAL AND MICROSCOPIC EXAMINATION OF THE BALLOON OBSERVED A LESS THAN 1MM LONGITUDINAL TEAR IN THE BALLOON MATERIAL APPROXIMATELY 22MM PROXIMAL TO THE PROXIMAL EDGE OF THE DISTAL MARKERBAND. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE NON-CALCIFIED CEPHALIC VEIN. ON THE FIRST INFLATION OF THE 6.0X40MM WANDA PTA BALLOON IT WAS INFLATED TO 12 ATM WITH AN ENCORE 26 INFLATION DEVICE AND THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE NON-CALCIFIED CEPHALIC VEIN. ON THE FIRST INFLATION OF THE 6.0X40MM WANDA PTA BALLOON IT WAS INFLATED TO 12 ATM WITH AN ENCORE 26 INFLATION DEVICE AND THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WANDA PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - GALWAY H965SCH505100 14066228

Patients

Seq Age Sex Outcome Treatment
1 60 YR