PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2133696 · Received June 9, 2011

Report

Report Number
3004209178-2011-81700
Event Type
Injury
Date Received
June 9, 2011
Date of Event
May 23, 2011
Report Date
May 23, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HYPERGLYCEMIA, WITH BLOOD GLUCOSE LEVELS OVER 600 MG/DL. THE CUSTOMER CHANGED THE INFUSION SET THE DAY PRIOR TO THE EVENT. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. FOUND MULTIPLE NO DELIVERY ALARMS IN THE ALARM HISTORY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THERE WAS NO TUBING CLAMP FOR THE HIGH PRESSURE TEST. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 50 YR Hospitalization