FDA Adverse Event Injury Summary report: N

ANGIOSCULPT PTCA SCORING BALLOON CATHETER (EX)

MDR report key: 2133648 · Received June 8, 2011

Report

Report Number
3005462046-2011-00023
Event Type
Injury
Date Received
June 8, 2011
Date of Event
May 3, 2011
Report Date
May 12, 2011
Manufacturer
ANGIOSCORE, INC.
Product Code
NWX
PMA / PMN Number
P050018
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PROVIDED IS THE EVENT DESCRIPTION AS REPORTED IN THE CATHETERIZATION REPORT, WHICH WAS PROVIDED BY THE HOSPITAL. THE PROCEDURAL ANGIOGRAM WAS ALSO REQUESTED BUT NOT PROVIDED BY THE HOSPITAL. METHOD: PRODUCT DISPOSED BY HOSPITAL, THUS UNABLE TO EVALUATE DEVICE. RESULTS: CORONARY ARTERY DISSECTION IS LISTED AS A POSSIBLE ADVERSE EFFECT OF THE PROCEDURE IN THE ANGIOSCULPT DEVICE INSTRUCTIONS FOR USE (IFU).

Description of Event or Problem · 1

A 2.0 X 10 ANGIOSCULPT WAS USED TO SCORE/PRE-DILATE THE OSTIUM OF OM2. THE CLINICIAN INFLATED THE ANGIOSCULPT SLOWLY, PER THE MANUFACTURER'S RECOMMENDATIONS, FOR APPROXIMATELY A TOTAL OF ONE (1) MINUTE. POST-SCORING ANGIOGRAPHY REVEALED A DISSECTION OF THE OSTIAL OM2. AS A RESULT OF THIS, THE CLINICIAN DECIDED TO STENT INTO THE OM2 ACROSS THE TRUE AV GROOVE CIRCUMFLEX. THE OSTIUM OF THE TRUE AV GROOVE CIRCUMFLEX WAS PRE-DILATED WITH A 3.0 X 12 COMPLIANT BALLOON WITH EXCELLENT RESULTS. A 3.0 X 22 INTEGRITY BARE-METAL STENT WAS DEPLOYED AT 14 ATMOSPHERES FOR 30 SECONDS. THE FINAL ANGIOGRAPHIC RESULTS WERE EXCELLENT. PT IS CLINICALLY STABLE. NO ANGIOGRAPHIC COMPLICATIONS. SUCCESSFUL PCI OF COMPLICATED OM/LEFT CIRCUMFLEX BIFURCATION WITH STENT PLACEMENT (3.0 X 22 INTEGRITY BARE-METAL) IN THE OM2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANGIOSCULPT PTCA SCORING BALLOON CATHETER (EX) NWX ANGIOSCORE, INC. 2034-2010 F11010031

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention WIRE| ABBOTT HI-TORQUE BALANCE MIDDLEWEIGHT (BMW) GUIDE| ABBOTT ASAHI PROWATER CORONARY GUIDE WIRE