COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-08059
- Event Type
- Malfunction
- Date Received
- June 20, 2011
- Date of Event
- May 31, 2011
- Report Date
- June 13, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE DEPLETED MAIN BATTERIES AS A RESULT OF USER ERROR. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. A DEVICE HISTORY REVIEW REVEALED NO ABNORMALITIES WERE FOUND DURING MANUFACTURING. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO INFORMATION REGARDING PATIENT INVOLVEMENT, PATIENT INJURY, OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |