FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20E DEFIBRILLATOR/MONITOR

MDR report key: 2133245 · Received May 11, 2011

Report

Report Number
3015876-2011-00380
Event Type
Malfunction
Date Received
May 11, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K073089
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE THERAPY PCB AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE POWER BUTTON WAS PRESSED, THE DISPLAY WOULD FLASH BRIEFLY, BUT THE DEVICE WOULD NOT COMPLETELY POWER ON. THIS OCCURRED WITH BOTH AC AND DC POWER. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20E DEFIBRILLATOR/MONITOR MKJ PHYSIO-CONTROL, INC. 20E NA

Patients

Seq Age Sex Outcome Treatment
1 NA