FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 SLIDEMAKER ANALYZER

MDR report key: 2133177 · Received June 20, 2011

Report

Report Number
1061932-2011-00710
Event Type
Malfunction
Date Received
June 20, 2011
Date of Event
May 20, 2011
Report Date
May 20, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
LOQ
PMA / PMN Number
K011342
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND FOUND THE LEAK AT THE TUBING ON THE PERISTALTIC PUMP 1 AND REPLACED THE BATH 1 PERISTALTIC PUMP AND TUBING. THE FSE VERIFIED THE INSTRUMENTS OPERATION. THE ROOT CAUSE WAS ATTRIBUTED TO A LEAK FROM THE TUBING ASSOCIATED WITH THE BATH 1 PERISTALTIC PUMP. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE CUSTOMER FOUND FLUID IN THE TUBING TRAY ON THE COULTER LH 750 SLIDEMAKER ANALYZER. PER THE CUSTOMER, THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT AND APPEARED TO BE METHANOL REAGENT FROM THE SPLIT TUBING ON THE PUMP 1 FOR BATH 1. THE CUSTOMER DOES NOT HAVE A SPARE PUMP AND REQUESTED INSTRUMENT SERVICE. THE CUSTOMER WAS WEARING GLOVES AND A LAB COAT. NO INJURIES OCCURRED AND MEDICAL ATTENTION WAS NOT SOUGHT. THERE WAS NO REPORT OF EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THERE WAS NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED WITH THIS COMPLAINT.