FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 2133167 · Received June 19, 2011

Report

Report Number
2122870-2011-01817
Event Type
Malfunction
Date Received
June 19, 2011
Date of Event
May 18, 2011
Report Date
May 18, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K922823/A007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULT: ERRONEOUS DATA GENERATED. CONCLUSION: A DEFINITIVE ROOT CAUSE HAS NOT YET BEEN DETERMINED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) TO REPORT A LOWER THAN EXPECTED RESULT FOR FERRITIN FOR 1 PATIENT SAMPLE ASSAYED WITH ACCESS FERRITIN REAGENT ON AN ACCESS 2 IMMUNOASSAY SYSTEM. THE PATIENT RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT. QUALITY CONTROL RESULTS WERE WITHIN ESTABLISHED RANGES AT THE TIME OF THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE ON (B)(4) 2011. THE FSE REPLACED THE PERI-PUMP TUBING, ASPIRATE PROBES, SUBSTRATE PROBE, AND THE SAMPLE PIPETTOR. ALL VERIFICATION TESTING PERFORMED MET ANALYZER SPECIFICATIONS. A DEFINITIVE ROOT CAUSE HAS NOT YET BEEN DETERMINED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS® 2 IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER JJE BECKMAN COULTER, INC.

Patients

Seq Age Sex Outcome Treatment
1 ACCESS FERRITIN REAGENT