HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-07796
- Event Type
- Malfunction
- Date Received
- June 19, 2011
- Date of Event
- April 12, 2011
- Report Date
- June 10, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4).
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RECEIVED FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE OF THE IIPV FOUND DURING LOG REVIEW WAS DETERMINED TO BE INSUFFICIENT DRAIN-USE ERROR; THE TIDAL UF REMOVAL WAS SET TOO LOW (AT 0ML). A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE SUFFICIENT. A SERVICE HISTORY REVIEW (SHR) REVEALED THAT NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. ADDITIONAL INVESTIGATION IS BEING CONDUCTED THROUGH (B)(4).
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN THE LOGS THAT OCCURRED ON (B)(6) 2011 STARTED AT 02:56 WITH UF VOLUME OF 1354 ML DURING CYCLE 4. THE FILL VOLUME WAS 1500 ML. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.