FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2133081
·
Received June 18, 2011
Report
- Report Number
- 2531779-2011-04266
- Event Type
- Malfunction
- Date Received
- June 18, 2011
- Date of Event
- May 13, 2011
- Report Date
- May 20, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(6). THERE IS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING TESTING, ALL KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONSIVE. THE RUBBER KEYPAD APPEARED WORN BUT WAS NOT TORN OR PEELING. WHEN THE KEYPAD WAS REMOVED, ADHESIVE WAS FOUND UNDER THE KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTONS REQUIRE MULTIPLE PRESSES TO ELICIT THE INTENDED RESPONSE. THERE WAS NO REPORT OF PROGRAMMING ERRORS AS A RESULT OF THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |