ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-04264
- Event Type
- Malfunction
- Date Received
- June 18, 2011
- Date of Event
- May 18, 2011
- Report Date
- May 19, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
IT WAS REPORTED THAT THE KEYPAD BUTTONS REQUIRED MULTIPLE PRESSES TO ELICIT THE INTENDED RESPONSE. THERE WAS NO REPORT OF PROGRAMMING ERRORS AS A RESULT OF THIS ISSUE.