FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 2133038 · Received June 17, 2011

Report

Report Number
2955842-2011-00187
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 19, 2011
Report Date
May 19, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THE TUBE EXTENSION TO BE BROKEN AND MISSING A .280 X .210 PIECE AT THE PROX CLEVIS INTERFACE. THE SCISSOR BLADES ARE INTACT AND THE CLEVIS IS DISLODGED FROM TUBE EXTENSION. EVIDENCE NOT CONCLUSIVE, BUT TUBE EXTENSION LIKELY BROKE DUE TO EXCESSIVE SIDE LOADING. ELECTRICAL CONTINUITY PASSED. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, THE SITE OBSERVED THE TIP OF THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT TO HAVE BROKEN. THE PIECE WAS IMMEDIATELY RETRIEVED AND THE INSTRUMENT WAS REPLACED BEFORE THE PLANNED SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT INCIDENT. THE PATIENT WAS REPORTED AS DOING WELL WITH NO POST OPERATIVE COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM, INSTRUMENTS, ACCESSORIES, & ES