FDA Adverse Event Injury Summary report: N

SELECT SECURE

MDR report key: 2132928 · Received June 17, 2011

Report

Report Number
2649622-2011-10736
Event Type
Injury
Date Received
June 17, 2011
Date of Event
April 21, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P030036
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD BECAME DISLODGED AND TRACKED BACK THROUGH THE RIGHT ATRIAL SUTURE SLEEVE. THE LEAD WAS COILED BEHIND THE DEVICE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELECT SECURE IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 3830 ASKU

Patients

Seq Age Sex Outcome Treatment
1 11 YR Hospitalization| R 3830 IMPLANTABLE PACING LEAD| ADDRL1 IMPLANTABLE PULSE GENERATOR| ADDRL1 IMPLANTABLE PULSE GENERATOR| 3830 IMPLANTABLE PACING LEAD