COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-07949
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- May 22, 2011
- Report Date
- May 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-08/04/10-002-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE CURRENT USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.90, CATEGORIZED AS REMEDIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |