FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT

MDR report key: 2132732 · Received June 17, 2011

Report

Report Number
3005075853-2011-02459
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 25, 2011
Report Date
May 27, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION:THE STAPLES WERE DEPLOYED TILL THE CUT END OF THE STAPLE LINE. SOME DEPLOYED STAPLES OF THE ACRAL PART WERE MALFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN GASTRECTOMY PROCEDURE, THE KNIFE STOPPED ON THE WAY OF FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION IN H10.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE