PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT
Report
- Report Number
- 3005075853-2011-02459
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- May 25, 2011
- Report Date
- May 27, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION:THE STAPLES WERE DEPLOYED TILL THE CUT END OF THE STAPLE LINE. SOME DEPLOYED STAPLES OF THE ACRAL PART WERE MALFORMED.
IT WAS REPORTED THAT DURING AN OPEN GASTRECTOMY PROCEDURE, THE KNIFE STOPPED ON THE WAY OF FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION IN H10.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |