FDA Adverse Event Injury Summary report: N

MINICAP TRANSFER SET

MDR report key: 2132721 · Received June 17, 2011

Report

Report Number
1423500-2011-07760
Event Type
Injury
Date Received
June 17, 2011
Date of Event
April 1, 2011
Report Date
May 25, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS REPORT 4 OF 4 INVOLVED IN THIS PERITONITIS EVENT. THE SAMPLE IS NOT AVAILABLE. A FOLLOW -UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION OF PERITONITIS IS USE ERROR- POOR ASEPTIC TECHNIQUE. A BATCH REVIEW WAS PERFORMED FOR THE POTENTIALLY ASSOCIATED LOT (H10H23052). THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. A LABELING REVIEW WAS PERFORMED AND THE LABELING WAS FOUND TO PROVIDE AMPLE INSTRUCTIONS RELATED TO PREVENTION OF THE USE ERROR IN ASEPTIC TECHNIQUE. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE. THIS IS A SPONTANEOUS REPORT BY A NURSE IN THE USA OF PERITONITIS COINCIDENT WITH DIANEAL PD4 AMBUFLEX THERAPY. ON AN UNREPORTED DATE, THE PATIENT BEGAN TREATMENT WITH DIANEAL PD4 AMBUFLEX (DOSE AND FREQUENCY NOT REPORTED) INTRAPERITONEALLY (IP) FOR PERITONEAL DIALYSIS (PD). DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE NURSE STATED THAT ON (B)(6) 2011, THE PATIENT EXPERIENCED PERITONITIS AND WAS HOSPITALIZED THAT SAME DAY. THE CAUSE OF THE PERITONITIS WAS UNKNOWN. THE NURSE STATED THAT THE CAUSE WAS POSSIBLY DUE TO A BREAK IN ASEPTIC TECHNIQUE. TREATMENT INFORMATION WAS NOT REPORTED. IN (B)(6) 2011, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL. ON AN UNREPORTED DATE, DIANEAL THERAPY WAS WITHDRAWN. IN 2011, THE PATIENT RECOVERED. THE NURSE STATED THAT THE EVENT OF PERITONITIS WAS UNRELATED TO DIANEAL THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 59 YR HOMECHOICE| DIANEAL PD4 AMBUFLEX