FDA Adverse Event Malfunction Summary report: N

EQUALIZER BALLOON CATHETER

MDR report key: 2132720 · Received June 17, 2011

Report

Report Number
2134265-2011-02308
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 24, 2011
Report Date
May 24, 2011
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
DQY
PMA / PMN Number
K021721
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING PROCEDURE, A BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE MODERATELY TORTUOUS TARGET LESION WAS LOCATED IN THE AORTA. THE EQUALIZER (B)(6) BALLOON WAS USED FOR POST-DILATATION AND RUPTURED UPON THE 1ST INFLATION AT AN UNKNOWN ATM. THE DEVICE WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EQUALIZER BALLOON CATHETER CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - CORK M001171120 13847352