HARMONIC ACE CURVED SHEARS INSERT
Report
- Report Number
- 2955842-2011-00179
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- April 26, 2011
- Report Date
- May 18, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K093217
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- NURSE
Narratives
THE HARMONIC ACE INSERT ACCESSORY HAS NOT BEEN RETURNED FOR EVALUATION, HOWEVER, AN ISI PRODUCT MANAGER PERFORMED A SITE VISIT ON (B)(6) 2011 TO OBSERVE CASES, SURGEON AND SURGICAL STAFF TECHNIQUE, AND PROVIDE FEEDBACK TO THE SITE REGARDING FREQUENT HARMONIC ACE INSERT BREAKAGE. IT WAS OBSERVED THAT THE BREAKAGE THE SITE IS EXPERIENCING IS OCCURRING WHEN THE ACTIVE BLADE COMES IN CONTACT WITH METAL OR THE BLADE IS DAMAGED DUE TO INTRA OPERATIVE COLLISIONS OR OTHER PHYSICAL CONTACT WITH OTHER INSTRUMENTS. THE DA VINCI HARMONIC ACE CURVED SHEARS INSTRUCTIONS FOR USE SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: AVOID CONTACT WITH ANY AND ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE DEVICE IS ACTIVATED. CONTACT WITH STAPLES, CLIPS, OR OTHER INSTRUMENTS WHILE THE DEVICE IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY A GENERATOR SOLID TONE OR AN INSTRUMENT ERROR. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED DUE TO THE HARMONIC ACE INSERT ACCESSORY NOT BEING RETURNED FOR EVALUATION. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, WHEN THE SURGEON BEGAN TO AMPUTATE THE UTERUS FROM THE CERVIX, THE TIP OF THE HARMONIC ACE CURVED SHEARS INSERT BROKE AND A PIECE FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE CURVED SHEARS INSERT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400272 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYS, INSTRUMENTS, ACCESSORIES, & ESU |