FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750

MDR report key: 2132519 · Received June 17, 2011

Report

Report Number
1061932-2011-00664
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 18, 2011
Report Date
May 18, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT IS CURRENTLY PERFORMING WITH QC SPECIFICATIONS WITH RESPECT TO CONTROLS AND REPRODUCIBILITY. QC WAS DONE BEFORE AND AFTER THE EVENT. A FIELD SERVICE ENGINEER (FSE) REPLACED VC37 AND THE ISSUE WAS RESOLVED. THE ROOT CAUSE IS ATTRIBUTED TO VACUUM CHAMBER 37 LEAKING AIR AND NOT SUPPLYING PRESSURE TO THE BACKWASH TANK, WHICH IN TURN SUPPLIES DILUENT TO THE HGB CUVETTE FOR THE BLANK READING.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) AND REPORTED THAT THE ERRONEOUS HEMOGLOBIN (HGB) RESULTS WERE GENERATED BY THE COULTER LH 750 ANALYZER FOR THREE PATIENTS WITH NO INSTRUMENT GENERATED FLAGS. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LAB, AND WERE ONLY DETECTED AFTER FAILING BACKGROUND CHECKS AND CONTROLS FAILING QC AT THE START OF ANOTHER SHIFT. THE SAMPLES WERE RE-TESTED ON ANOTHER INSTRUMENT AND THE RESULTS OBTAINED WERE THE CORRECT RESULTS AND WERE REPORTED OUT OF THE LAB. THE CUSTOMER CONFIRMED THAT THERE WAS NO DEATH OR CHANGE TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 750 HEMATOLOGY ANALYZER GKZ BECKMAN COULTER, INC. LH 750 NA

Patients

Seq Age Sex Outcome Treatment
1