CAVITRON SELECT SPS G124 GENERATOR(DNA)
Report
- Report Number
- 2424472-2025-00017
- Event Type
- Malfunction
- Date Received
- February 6, 2025
- Report Date
- April 7, 2025
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- UDI-DI
- D003813021
- PMA / PMN Number
- K970123
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
(B)(6) 2025 REPAIR TECH: MTK. MPH CAB, COVER, LID, LID HINGES DAMAGED. FOUND THAT THE RESERVOIR FOR THE PUMP IS CRACKED WHERE THE CAP SCREWS ON REPLACED DAMAGED/WORN COMPONENTS AND RECALIBRATED UNIT TO FACTORY SPECS. QA: DS.
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE LACK OF WATER FLOW HAS CAUSED AN OVERHEATING INSERT. SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS OVERHEATING; NO INJURY RESULTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1106789 | CAVITRON SELECT SPS G124 GENERATOR(DNA) | SCALER, ULTRASONIC | ELC | DENTSPLY LLC | D003813021 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |