FDA Adverse Event
Malfunction
Summary report: N
SYSTEM 6 CHARGER
MDR report key: 2132328
·
Received June 11, 2011
Report
- Report Number
- 1811755-2011-02137
- Event Type
- Malfunction
- Date Received
- June 11, 2011
- Date of Event
- May 19, 2011
- Report Date
- May 19, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MOST LIKELY CAUSE OF THE FAILURE WAS THE PROTECTION CIRCUIT WITHIN THE CHARGER POTENTIALLY DETERMINED THE POWER SUPPLY WAS UNABLE TO SAFELY MEET THE CURRENT DRAW AND SHUT DOWN THE CHARGING CIRCUITRY. THE BATTERIES WERE SUBJECTED TO EXTENDED EXPOSURE TO HIGH VOLTAGE, FROM THE BATTERY CHARGER RESETTING CONTINUOUSLY BECAUSE ALL FOUR BATTERIES WERE INITIALIZING AT THE SAME TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE BATTERIES MELTED WHILE BEING CHARGED. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYSTEM 6 CHARGER | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |