FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 2132321 · Received June 11, 2011

Report

Report Number
1811755-2011-02089
Event Type
Malfunction
Date Received
June 11, 2011
Date of Event
May 23, 2011
Report Date
May 23, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. THE DEVICE EVALUATION REVEALED THAT THE DRILL PERFORMED ACCORDING TO ALL SPECIFICATIONS, HOWEVER, A HOLE IN THE PNEUMATIC HOSE ASSEMBLY WAS CONFIRMED. IT IS UNKNOWN HOW THE HOSE DAMAGE OCCURRED, BUT MAY BE THE RESULT OF MISHANDLING. THE HOSE ASSEMBLY WAS REPLACED AND THE DRILL WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOLE WAS DISCOVERED IN THE HOSE PORTION OF THE PNEUMATIC DRILL. THERE WAS NO REPORT OF ANY OIL LEAKAGE FROM THE DEVICE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK