FDA Adverse Event
Malfunction
Summary report: N
MAESTRO DRILL
MDR report key: 2132321
·
Received June 11, 2011
Report
- Report Number
- 1811755-2011-02089
- Event Type
- Malfunction
- Date Received
- June 11, 2011
- Date of Event
- May 23, 2011
- Report Date
- May 23, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RECEIVED BY THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. THE DEVICE EVALUATION REVEALED THAT THE DRILL PERFORMED ACCORDING TO ALL SPECIFICATIONS, HOWEVER, A HOLE IN THE PNEUMATIC HOSE ASSEMBLY WAS CONFIRMED. IT IS UNKNOWN HOW THE HOSE DAMAGE OCCURRED, BUT MAY BE THE RESULT OF MISHANDLING. THE HOSE ASSEMBLY WAS REPLACED AND THE DRILL WAS RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT A HOLE WAS DISCOVERED IN THE HOSE PORTION OF THE PNEUMATIC DRILL. THERE WAS NO REPORT OF ANY OIL LEAKAGE FROM THE DEVICE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO DRILL | HBB | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |