FDA Adverse Event Malfunction Summary report: N

UNIVERSAL HIGH TORQUE DRILL

MDR report key: 2132290 · Received June 11, 2011

Report

Report Number
1811755-2011-02117
Event Type
Malfunction
Date Received
June 11, 2011
Date of Event
March 28, 2011
Report Date
May 12, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY THE MFR AND THE EVENT WAS CONFIRMED. THERE WAS CORROSION DAMAGE ON THE MOTOR, ROTOR ASSEMBLY, AND BEARINGS. THE MOTOR WAS FOUND TO HAVE SEIZED DUE TO THE CORROSION DAMAGE TO THE ROTOR ASSEMBLY, BEARINGS AND ATTACHMENT LOCK. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. THERE WERE NO RELEVANT NON-CONFORMANCES, ALERTS, DEVIATIONS, OR REWORK RELATED TO THIS CONFIRMED FAILURE. THERE WERE NO RELEVANT DESIGN/PROCESS CHANGES RELATED TO THIS CONFIRMED FAILURE.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS RUNNING AT HIGH TEMPERATURES. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIVERSAL HIGH TORQUE DRILL HBE STRYKER INSTRUMENTS KALAMAZOO 0523500683

Patients

Seq Age Sex Outcome Treatment
1 UNK