FDA Adverse Event
Malfunction
Summary report: N
UNIVERSAL HIGH TORQUE DRILL
MDR report key: 2132290
·
Received June 11, 2011
Report
- Report Number
- 1811755-2011-02117
- Event Type
- Malfunction
- Date Received
- June 11, 2011
- Date of Event
- March 28, 2011
- Report Date
- May 12, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY THE MFR AND THE EVENT WAS CONFIRMED. THERE WAS CORROSION DAMAGE ON THE MOTOR, ROTOR ASSEMBLY, AND BEARINGS. THE MOTOR WAS FOUND TO HAVE SEIZED DUE TO THE CORROSION DAMAGE TO THE ROTOR ASSEMBLY, BEARINGS AND ATTACHMENT LOCK. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. THERE WERE NO RELEVANT NON-CONFORMANCES, ALERTS, DEVIATIONS, OR REWORK RELATED TO THIS CONFIRMED FAILURE. THERE WERE NO RELEVANT DESIGN/PROCESS CHANGES RELATED TO THIS CONFIRMED FAILURE.
Description of Event or Problem · 1
IT WAS REPORTED THE DEVICE WAS RUNNING AT HIGH TEMPERATURES. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL HIGH TORQUE DRILL | HBE | STRYKER INSTRUMENTS KALAMAZOO | 0523500683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |