FDA Adverse Event Injury Summary report: N

SHOULDER SYSTEM

MDR report key: 21321671 · Received February 6, 2025

Report

Report Number
3005180920-2025-00041
Event Type
Injury
Date Received
February 6, 2025
Date of Event
January 14, 2025
Report Date
February 6, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706292
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 16-01-2024. LOT 2117046: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-01-2022. EXPIRATION DATE: 2027-01-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL IMPLANT INVOLVED; BATCH REVIEW PERFORMED ON 16-01-2024. REVERSE SHOULDER SYSTEM 04.01.0170 GLENOSPHERE 39XØ24.5 (K170452) LOT 1910953: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-06-2020. EXPIRATION DATE: 2025-05-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

ON (B)(6) 2018 THE PATIENT HAD A PRIMARY ANATOMIC SHOULDER SURGERY. ON (B)(6) 2024, THE SURGEON CONVERTED THE PATIENT FROM AN ANATOMICAL SHOULDER TO A REVERSE SHOULDER DUE TO THE PATIENT REPORTING PAIN RELATED TO A ROTATOR CUFF FAILURE. ON (B)(6) 2024, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE GLENOSPHERE FROM THE LINER. THE SURGEON CLOSED REDUCED UNDER ANESTHESIA THE PATIENT AND NO IMPLANTS WERE REVISED. THE CASE WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2025, THE PATIENT REPORTED PAIN DUE TO A DISLOCATION OF THE LINER FOR THE GLENOSPHERE AFTER LIFTING HEAVY ITEMS. THE SURGEON REVISED THE LINER, GLENOSPHERE AND METAPHYSIS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
919899 SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø39/+0MM PHX MEDACTA INTERNATIONAL SA 04.01.0122 2117046 07630040706292

Patients

Seq Age Sex Outcome Treatment
1 79 YR Male Required Intervention