FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2132135 · Received June 17, 2011

Report

Report Number
6000001-2011-07904
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
June 1, 2011
Report Date
June 1, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS CONFIRMED. UPON SAMPLE RECEIPT, THE DEVICE CONTAINED FLUID IN THE BLADDER. TO VERIFY THE REPORTED CONDITION, THE FILL-PORT CAP WAS REMOVED. UPON REMOVAL, A BACK-FLOW (LEAK) WAS OBSERVED AT THE FILL-PORT. VISUAL EXAMINATION OF THE DISASSEMBLED UNIT REVEALED THE ROOT CAUSE OF THE LEAK WAS A 1.1-MM PARTICLE OF ACRYLIC RESIN TRAPPED BETWEEN THE CHECK BAND AND THE STRESS MEMBER. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT WITH ONE (1) INFUSOR, DURING FILLING, BACKFLOW WAS OBSERVED FROM THE FILLING PORT. THE DEVICE WAS BEING FILLED WITH SALINE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THERE WAS NO PATIENT INVOLVEMENT. THE ACTUAL SAMPLE IS AVAILABLE. NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 11B085

Patients

Seq Age Sex Outcome Treatment
1 SALINE