ACCESS 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-01908
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- April 29, 2011
- Report Date
- May 17, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- 922823/A007
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
SAMPLE COLLECTION DEVICE AND CENTRIFUGE INFORMATION HAVE NOT BEEN SUPPLIED TO DATE. SYSTEM CHECK COMPLETED ON (B)(6) 2011 PASSED WITHIN INSTRUMENT SPECIFICATIONS. QC PERFORMED WITHIN THE CUSTOMER'S SPECIFICATIONS. SERVICE INFORMATION HAS NOT BEEN PROVIDED. THE SAMPLE WAS SENT TO CUSTOMER PRODUCT LINE SUPPORT (CPLS) FOR INVESTIGATIONAL TESTING. (B)(6) RESULT OBTAINED BY THE LAB ON THIS SAMPLE WAS CONFIRMED BY THE CPLS LAB. HETEROPHILE TESTING ON IGM AND IGG ELUTION FRACTIONS DID NOT DEMONSTRATE THE PRESENCE OF AN INTERFERING COMPOUND. CPLS DID NOT FIND EVIDENCE OF A DEFECTIVE REAGENT. A DEFINITIVE ROOT CAUSE CAN NOT BE DETERMINED FOR THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS (B)(6) HEPATITIS B ANTIBODY RESULTS THAT DID NOT FIT THE CLINICAL PICTURE FOR ONE PATIENT GENERATED BY THE ACCESS 2 IMMUNOASSAY ANALYZER. THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THE INITIAL SAMPLE WAS RERUN ON THE SAME UNIT AND ON AN ALTERNATE METHOD. THE REPEAT RESULTS ON ACCESS TWO MATCHED THE INITIAL RUN RESULTS. THE ALTERNATE METHOD RESULT DID NOT MATCH THE (B)(6) HEPATITIS B ANTIBODY RESULTS OBTAINED ON THE ACCESS 2 IMMUNOASSAY ANALYZER. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | ACCESS 2 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |