FDA Adverse Event Malfunction Summary report: N

ASKU

MDR report key: 2132069 · Received June 16, 2011

Report

Report Number
2183613-2011-00261
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED EVENT. THE LCD DISPLAY IS OUT OF SPECIFICATION. THE UPPER AND LOWER CASES AND THE BATTERY DRAWER FOUND BROKEN. THE RING IS BENT.

Description of Event or Problem · 1

IT WAS REPORTED THE RIGHT SIDE OF THE LOWER DISPLAY HAD A VERTICAL LINE OF MISSING PIXELS. THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other