FDA Adverse Event Malfunction Summary report: N

ASKU

MDR report key: 2132012 · Received June 16, 2011

Report

Report Number
2183613-2011-00225
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE DEVICE WOULD NOT POWER UP. THE CAUSE WAS THE MAIN PCB (PRINTED CIRCUIT BOARD). THE BATTERY CONTACTS WERE ALSO FOUND TO BE COMPRESSED, THE BATTERY RELEASE AND LEAD FLEX COVER WERE CONTAMINATED, AND THE RING COVER, TWO SIDE BAIL COVERS, AND UPPER AND LOWER CASES WERE BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT WILL NOT TURN ON. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other