MDR report key: 2132011 · Received June 16, 2011

Report

Report Number
2182208-2011-00947
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE DEVICE WOULD NOT POWER UP. THE CAUSE WAS THE MAIN PCB (PRINTED CIRCUIT BOARD). THE LOWER CASE WAS ALSO NOTED TO BE BROKEN AND THE KEYBOARD PAD SCRATCHED.

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT WOULDN'T TURN ON, AND IT WAS POWERING DOWN ON ITS OWN. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. FOLLOW-UP INFORMATION RECEIVED DETERMINED THERE WAS NO PATIENT INVOLVEMENT. THE CONDITION WAS FOUND PRIOR TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC, INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other