FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2131975 · Received June 16, 2011

Report

Report Number
2024168-2011-04282
Event Type
Injury
Date Received
June 16, 2011
Date of Event
May 25, 2011
Report Date
May 26, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. WITHOUT THE DEVICE, A CAUSE FOR THE SUTURE BREAK COULD NOT BE DETERMINED. A SUTURE BREAK CAN BE INFLUENCED BY MANY FACTORS INCLUDING, BUT NOT LIMITED TO, MANUFACTURING, TOO MUCH TENSION APPLIED, OR THE SUTURE WAS NICKED. WHETHER THESE CONDITIONS PLAYED A ROLE IN THE EXPERIENCED EVENT CANNOT BE VERIFIED. TO ASSURE THAT ALL PRODUCTS PERFORM ACCORDING TO SPECIFICATIONS A SAMPLING OF SUTURES ARE TESTED UNDER STRESS FOR DIAMETER MEASUREMENTS BEFORE RELEASE TO MANUFACTURING. IN ADDITION, A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. THE LOT HISTORY RECORD FOR THIS PRODUCT WAS NOT REVIEWED AND A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS NOT PERFORMED AS THE LOT NUMBER WAS NOT REPORTED AND THE PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION RECEIVED WITH THIS EVENT, THERE DOES NOT APPEAR TO BE ANY INDICATION OF AN ISSUE WITH THE QUALITY OF THE PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A TECHNICIAN, TRAINED IN THE USE OF THE PERCLOSE PROGLIDE, ATTEMPTED ARTERIOTOMY CLOSURE OF THE MILDLY CALCIFIED COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, DURING KNOT ADVANCEMENT, THE SUTURE BROKE. MANUAL ARTERIAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT

Patients

Seq Age Sex Outcome Treatment
1 43 YR Required Intervention