SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-10617
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THE DEFIB CONDUCTOR WAS DISTORTED AND THE INNER TUBING WAS KINKED/BUCKLED. THE OUTER TUBING OVERLAY WAS MELTED AND HAD COSMETIC ENVIRONMENTAL STRESS CRACKING. SEVERAL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED) AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVE HEAD. THERE WAS A TIP SEAL OBSERVATION AND APPARENT EXPLANT DAMAGE.
ASKU
IT WAS REPORTED THAT THE LEAD HAD HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| R | 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD |