FDA Adverse Event Injury Summary report: N

OT ULTRA METER

MDR report key: 2131929 · Received June 16, 2011

Report

Report Number
2939301-2011-04920
Event Type
Injury
Date Received
June 16, 2011
Report Date
June 10, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K062195.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE PRODUCTS HAVE BEEN RETURNED TO LIFESCAN IN A CONDITION THAT MADE ANALYSIS IMPOSSIBLE BECAUSE TEST STRIPS WERE EXPIRED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(6) 2011) - DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A LOW/DEAD BATTERY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRA METER WOULD NOT POWER ON. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON (B)(6), 2011 AT 940AM. THE CCA WAS ADVISED THE PATIENT MANAGES HER DIABETES WITH METFORMIN PILLS (500 MG). THE PATIENT DENIED MAKING CHANGES TO HER DIABETES MANAGEMENT ROUTINE. ABOUT 10 HOURS AFTER THE ALLEGED ISSUE BEGUN, THE PATIENT CLAIMED SHE FELT A SYMPTOM OF DRY MOUTH. THE PATIENT DENIED RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THERE WAS NO MISUSE OF THE SUBJECT METER. THE CCA WAS UNABLE TO WALK THROUGH RESOLVING THE ALLEGED ISSUE. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMS SHE WAS UNABLE TO TEST HER BLOOD GLUCOSE DUE TO THE REPORTED ISSUE AND REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF HYPERGLYCEMIA AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRA METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 2629590

Patients

Seq Age Sex Outcome Treatment
1 52 YR Life Threatening