FDA Adverse Event Malfunction Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 2131904 · Received June 16, 2011

Report

Report Number
2955842-2011-00177
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
May 17, 2011
Report Date
May 17, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED TWO HOLES BURNED THOUGH THE SILICONE. THE SILICONE EXHIBITS LOCALIZED MELTING AROUND THE HOLE WHICH IS INDICATIVE OF AN ARCING EVENT. THE HOLE SIZES ARE .055 AND .045 X .075 WITH VARYING SHAPES (ROUND AND OBLONG), AND ARE LOCATED .075 AND .320 ABOVE THE SILICONE TO SLEEVE INTERFACE. AN ADDITIONAL INVESTIGATION IS BEING CONDUCTED TO DETERMINE ROOT CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP MDR WILL BE PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI S HYSTERECTOMY PROCEDURE, TWO HOLES WERE OBSERVED TO HAVE BEEN BURNED IN THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT TIP COVER ACCESSORY INSTALLED ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. NO ARCING WAS OBSERVED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180-10 C11038

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS, ACCESSORIES, & ESU