FDA Adverse Event
Injury
Summary report: N
CONSULTA CRT-D
MDR report key: 2131897
·
Received June 16, 2011
Report
- Report Number
- 6000144-2011-03504
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S084
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS NOT PACING AT A THE PROGRAMMED RATE. FOLLOW-UP DETERMINED THAT THE PATIENT HAD INTERMITTENT T-WAVE OVERSRENSING (TWOS) WHICH CAUSED PACING TO BE DELIVERED BELOW THE LOWER RATE. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONSULTA CRT-D | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | D224TRK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention | 4193 IMPLANTABLE PACING LEAD| 7121 COMPETITOR IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD |