FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2131897 · Received June 16, 2011

Report

Report Number
6000144-2011-03504
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS NOT PACING AT A THE PROGRAMMED RATE. FOLLOW-UP DETERMINED THAT THE PATIENT HAD INTERMITTENT T-WAVE OVERSRENSING (TWOS) WHICH CAUSED PACING TO BE DELIVERED BELOW THE LOWER RATE. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention 4193 IMPLANTABLE PACING LEAD| 7121 COMPETITOR IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD